Numéro
J. Phys. Radium
Volume 19, Numéro 3, mars 1958
Page(s) 256 - 259
DOI https://doi.org/10.1051/jphysrad:01958001903025600
J. Phys. Radium 19, 256-259 (1958)
DOI: 10.1051/jphysrad:01958001903025600

Emploi de fentes courbes pour la mise en série d'un réseau de diffraction avec un interféromètre fabry-Perot balayé par variation de pression

Joseph G. Hirschberg

The University of Wisconsin, Madison, Wisconsin, U. S. A.


Abstract
In the case of certain high resolution measurements, absorption for instance, the usual method, developed by P. Jacquinot and others, of employing the central region of the Fabry-Perot pattern may result in a loss of light due to geometry. For instance, with an f/10 grating with a 1 cm high exit slit, a 50 mm diameter Fabry-Perot, and a desired resolution of 5 x 105 in absorption, the light loss due to geometry is of the order of 4/5. It is shown that the situation can be much improved in certain cases by employing curved slits to match the curvature of the Fabry-Perot interference rings.


Résumé
Dans le cas de certaines mesures à haute résolution, en absorption par exemple, la méthode habituelle développée par P. Jacquinot et autres qui utilise la partie centrale de la figure de Fabry-Perot peut occasionner une perte de lumière en raison de la géométrie du système. Par exemple avec un réseau f/10 à fente de sortie de 1 cm de haut, un Fabry-Perot de 50 mm de diamètre et une résolution désirée de 5.105, en absorption, la perte de lumière due à des raisons géométriques est de l'ordre de 4/5. On montre que la situation peut, dans certains cas, être beaucoup améliorée par l'emploi de fentes courbes pour s'adapter à la courbure des anneaux d'interférence de Fabry-Perot.

PACS
4230 - Imaging and optical processing.

Key words
diffraction gratings -- light interferometers